ENGINEERING.com has updated it's forum.  To post a question please visit the new Ask@ Forum.   

With a database of over 10000 questions the library will remain available for an extended period.

Q&A


thegg1guy
HVAC Laboratory fume hood testing View All
Does anyone have information on Lab fume hood testing ?

17 years ago - 7 months left to answer. - 1 response - Report Abuse
Respond to question
    0      [lnkReport]        0       0       
Share |
  Responses


Gimpy
Fume Hoods are a primary engineering control used in chemical laboratories to provide personnel protection from chemical exposure. According to the Canadian Standards Association (CSA Standard Z316.5-94) fume hoods should be performance tested at least once per year. Some of these tests include:

1. Air Flow Measurement: Face velocity, Duct traverse, Smoke patterns. 2. Physical Characteristics: Damper/baffle operations, Sash operation, Drain operation, 3. Stability: Commissioning, Electrical safety, Lighting and 4. Noise: Vibration, Chemical Evaluation.


17 years ago

Source:


  1     0         

ENGINEERING.com does not provide engineering advice. The Ask@ service is a forum for members to exchange ideas relating to the world of engineering. We caution users not to accept any responses that they receive without further validation, and not to rely on any engineering advice that they may get from other members of the Ask@ forum. ENGINEERING.com specifically disclaims any obligation to validate or verify any information posted within the Ask@ service. ENGINEERING.com encourages users to seek the services of a professional engineer for any engineering advice they may require.